An artificial intelligence–enabled digital stethoscope demonstrates moderate murmur detection in dogs but not cats and unreliable arrhythmia classification in both species

Jake H. Johnson, Joshua A. Stern, Teresa C. DeFrancesco, Kursten V. Pierce

Journal of the American Veterinary Medical AssociationAug 5, 2026
Abstract

ObjectiveTo prospectively evaluate the diagnostic performance of an AI-enabled digital stethoscope in detecting cardiac murmurs and arrhythmias compared to fourth-year veterinary students and experienced clinicians.

MethodsDogs and cats presenting to a university teaching hospital were prospectively enrolled from August 1, 2025, through December 31, 2025. Each animal underwent cardiac auscultation at 4 thoracic sites with the use of an AI-enabled digital stethoscope (Core 500; EKO Health Inc), 6-lead ECG, and echocardiogram, if clinically indicated. Auscultation was performed by a cardiology resident, board-certified cardiologist, and fourth-year veterinary student.

ResultsThe stethoscope demonstrated a sensitivity of 86.8%, specificity of 56.3%, and positive predictive value of 82.5% for murmur detection in dogs. There was no difference between the agreement of the AI stethoscope or students with a clinician (κ = 0.447). In cats, sensitivity was markedly lower (9.1%), with only 2 of 22 murmurs detected (κ = 0.081). The stethoscope demonstrated high sensitivity for atrial fibrillation (100%), but classified no dog as arrhythmia-free. Murmur grade was the only significant predictor of stethoscope murmur diagnosis, and high-grade murmurs (≥ 3) had significantly greater odds of detection (OR, 15.11).

ConclusionsThe AI-enabled stethoscope demonstrated moderate murmur detection performance in dogs, comparable to fourth-year veterinary students, but performed poorly in cats. Arrhythmia classification was unreliable in both species. Clinical Relevance: AI-enabled stethoscopes represent an emerging tool in veterinary medicine, but clinical validation is necessary before routine adoption. This study provides prospective performance data across small animals and examiner levels, identifying meaningful limitations regarding interpretation of this technology.

1Authors

AuthorAffiliationh-indexCitations
Jake H. JohnsonNorth Carolina State University00
Joshua A. SternNorth Carolina State University303,628
Teresa C. DeFrancescoNorth Carolina State University322,970
Kursten V. PierceNorth Carolina State University6128

Showing the abstract — retrieve the full paper via the Exa API.

Powered by the Exa API

See something wrong or missing on this page?